Approval

Conformity MarkingCE (European Union)

Electrical

Mains Voltage100/115/230 V (+/- 10%), 50/60 Hz
Power Consumption~ 100
Control Voltage24 V DC

Dimensions and Weight

Width x Depth x Height

Weight

Cover Box340 x 320 x 305 mm~ 10 KG
Cryo Stage243 x 142 x 77 mm~ 1 KG
Cryo Transfer Shuttle393 x 120 x 119 mm~ 1 KG
LN2 Pump Controller220 x 354 x 132 mm~ 5 KG
LN2 Pump1170 x 92 x 92 mm~ 2 KG
LN2 Dewar (Empty)200 x 200 x 758 mm~ 10 KG
LN2 Dewar (Full)~ 30 KG

Operating Conditions

Min/Max

Optimal optical behavior

Room Temperature18 to 25 °C +/- 1 °C22 °C +/- 0,5 °C
Room Rel. Humidity< 40%< 20%
Storage Temperature5 to 35 °C21 to 24 °C
Storage HumidityMax. 50%Around 40%
Room VentilationRequired at all times during operation

Operating Temperatures

Cooling Temperature-195 °C
Heating RangeUp to 60 °C
Cooling Time20 Minutes (From +22 to -195 °C)
Cryo Stage Metal Shell Range5 to 45 °C (Avoids Icing)

Dewar

LN2 Dewar Volume25 l
Dewar Temperature-196 °C