DFC7000 T | DFC7000 GT | |||
Camera type | Digital color, high-sensitivity, cooled camera for fluorescence and brightfield microscopy | Digital monochrome, high-sensitivity, cooled camera dedicated for fluorescence applications | ||
Housing | Aluminum, Size 120 mm x 93 mm x 150 mm (L × W × H) , Weight 1,900 g | |||
Sensor | Sony® ICX674AQG CCD, Exview HAD II technology, quad-tap | Sony® ICX674AL CCD, Exview HAD II technology, quad-tap | ||
Shutter | Interline transfer progressive scan, global shutter | |||
Pixel | 1920 x 1440 (~ 2.8 MP); 4,54 μm x 4,54 μm pixel size | |||
Full well capacity | > 15.000 electrons (e-) typical | |||
Color filter | RGB Bayer mask | no | ||
Exposure time | 4 μsec - 200 sec* | |||
Bit depth | 8 bit / 12 bit | |||
A/D converter | 16 bit | |||
Cooling | One stage Peltier, stabelized | |||
Binning modes | 2 x 2 (color/ mono); 3 x 3 (color/ mono, speed optimized); 4 x 4 (color/ mono); 5 x 5 (color/ mono, speed optimized); color sensitive binning: R,G, or B pixel only | 2 x 2 (mono ); 3 x 3 (mono); 4 x 4 (mono); 5 x 5 (mono); | ||
Partial scan | Freely definable ROI (region of interest), combination with binning possible | |||
Dark noise | <0,05 e- / px / sec | |||
Read out noise | typ. 6 e- / 10 MHz | |||
Dynamic range | ~ 68 dB | |||
Pixel clocking rate | 10 MHz / 40 MHz | |||
Analog gain | continuous 1× – 10× | |||
Advanced features* | Image averaging, sharpening with active noise reduction, black balance, on-head image buffer, quad-tap read-out and mono-tap read-out, external trigger capability; dynamic defect pixel correction, HDR acquisition, streaming mode | |||
Image formats | fps*at 40 MHz / 8 bit | fps*at 40 MHz / 8 bit | ||
Turbo Mode (1280 x 1024) | 50 | 50 | ||
Full frame (1920 x 1440) | 40 | 40 | ||
2x2 binning (960 x 720) | 40 | 69 | ||
3x3 binning (640 x 480) | 91 | 91 | ||
4x4 binning (480 x 360) | 40 | 108 | ||
5x5 binning (384 x 288) | 123 | 124 | ||
Supported operating systems | Windows 7, Windows 8 (32 / 64 bit)*, Windows 10 | |||
Leica Software | LAS X, LAS | LAS X | ||
c-mount | 0.7x for inverted and upright compound microscopes; 0.63x for stereo microscopes | |||
Interfaces | USB 3.0 single cable (5 Gbit/s) and optional trigger cable | |||
Power supply | 5V via external DC power supply | |||
Power consumption | ~14 W (with Peltier cooling) | ~15 W (with Peltier cooling) | ||
Operating temp. range | +5°C to +50°C | |||
Storage temperature | -20° C to +70° C | |||
Air humidity | max 80%, non-condensing | |||
Conformity | CE: EN 61326-1, limited class B, EN 61010-1 ROHS 2 compliant, China ROHS 50 compliant |
(all information subject to modification without prior notice)
*depends on software in use